CONTRAINDICATIONS

Indications and Contraindications of QRS®-1010 PelviCenter
Electromagnetic Training of the muscles of the Pelvic Floor

Type: QRS® DTX 1010

Additional therapies: Oxygen therapy, QRS® relax therapy, Light therapy
INDICATIONS
⦁ Strengthening/ tightening of the muscles of the Pelvic Floor muscles
1 Recovery after pregnancy
2 Incontinence

⦁ Urinary Stress Incontinence
1 Urinary Urge Incontinence/ Overactive Bladder Syndrome
2 Mixed Incontinence
3 Fecal Incontinence
4 Early rehabilitation after radical prostatectomy
⦁ Erectile Dysfunction

CONTRA INDICATIONS ABSOLUTE

Pregnancy
Electronic Implants (Elektronische Implantate (Insulin / Pacemakers, Defibrallators)
Epileptic fits
Metal implants from Iron, Kobalt, Nickel, Gadolinium, Dysprosium, Alloys of Coppoer or Mangan. If closer than 20 cm from the seating of the PelviCenter the metals could get heated.
Recent surgery / recent radiation < 4 weeks ago
Severe heart rhythm disorders

CONTRA INDICATIONS RELATIVE

Alloys used in endo protesis (i.e. artificial hip) are normally not ferromagnetical (i.e. implants made of titanium) and are therefore SAFE Hip protesis close to the seating of the PelviCenter (could trigger muscle contractions)
Menstruation
Acute infections in Urinary organs/ urethra
Spirals
Painful hemorrhoids
Feverish infections
PATIENT SETTINGS
Empty bladder
Changing pads when wet
Removal of hearing aids/ hearing equipment
Removal of juwelry (bracelets, piercings, rings)
Removal of watches, care keys, credit cards, pagers, mobile phones, coins.